Assertion

AssertionId
DEC-Vol2-005 Testable
predicate
The PCD-01 message structure differs from the basic HL7 version 2.6 by allowing for the appearance of PRT segments, a segment new in HL7 version 2.7, in certain locations.
Prescription level
Permitted / May / Can
Page
14
Section
3.1.4.1.1 PCD-01 Communicate PCD Data (ORU^R01^ORU_R01) static definition
Status
to be reviewed
Last changed
2/26/16 10:23:33 AM by mpattillo
Comment

Covered by (Deprecated)

Tests

Test Steps

Rules

Document name
Provenance
Revision
Action
IHE Patient Care Device (PCD) Technical Framework - Volume 2IHE5.0