Special Instructions
In this test case, we will perform a RAD-69 transaction initiated by a Document Consumer simulator to the SUT acting as a XDS-I Document Source.
To perform this test, you'll need to register on your Document Source Repository the DICOM image files ("image_for_DOCUMENT_SOURCE.dcm") available at :
https://ehealthsuisse.ihe-europe.net/test_data/XDS-I_Documents/.
Patient to use : Nilesh WITTWER-CHRISTEN (EPR-SPID :
761337610411265304^^^SPID&2.16.756.5.30.1.127.3.10.3&ISO)
The Imaging Document Source actor is expected to act as a X-Service
provider. This test will also be used to assess this role for RAD-69
transaction. In order to do so, you'll need to follow the instructions
from this test case :
XUA_X-SERVICE-PROV_ITI-40
Important : TLS is mandatory to every request in the XDS profile, as a consequence every endpoint have to use HTTPS.
Description
Before anything, configure your system to be able to test its role as X-Service Provider, X-Service User and ADR Consumer.
The request will be sent to your SUT from Gazelle Webservice Tester. To execute this test:
- Access Gazelle Webservice Tester and log into the application;
- Go to menu "Run";
- Select the test project name EPR XDS-I Document Source from the drop-down list;
- Select test suite [RAD-69] Retrieve_Imaging_Document_Set TestSuite by ticking the checkbox in front of its name;
- Enter the URL of your system under test endpoint (shall be a secured endpoint)
- Click on "Run" button
- Wait for the script to complete its execution;
- When the script stops, copy the permanent link of the execution to the test step in Gazelle Test Management
A Passed result means the documents have been properly recovered by the simulator.
The
conformance of the responses sent back by your system shall be
assessed. In Gazelle Webservice Tester, validate the message in EVSClient. To do so:
- Click on the play icon next to the response type;
- You have been redirected to EVSClient, select XDS-CH in EPR group and click on the "Go" button;
- Once the page of the validator opens, select the appropriate entry in the drop-down list and click on "Validate";
- When the validation report shows up, a pop-up raises, click on "OK";
Evaluation
The monitor will check that you copied the link to this test instance to the corresponding test step of the test case XUA_X-SERVICE-PROV_ITI-40.
The validation on EVS Client must return Passed.
The status of your query must have a status "Passed" in GWT.