An analyzer supporting the "Query by Rack" Option SHALL be capable of sending a query identifying the position of the specimen in the rack (carrier). The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-4, QPD-5 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-4 identifies the rack and QPD-5 is the position in the rack
171
3.Q.5.4
VALIDATED
true