Assertion

AssertionId
DEC-Vol2-011 Testable
predicate
Pulse Oximetry Integration (POI) device observations shall include an observation of oxygen saturation of blood gas as a percentage numeric data type. Given the variety of available device designs (dedicated function SPO2 monitor or modular component to a physiologic monitor), the variety of sensor capabilities, and that the observation identification can include the sense point body part, i.e. left/right arm, etc. there is no single observation definition for all devices and all use cases. Any communicated SPO2 observation shall appear in RTMMS. A non-exclusive example is 150456^MDC_PULS_OXIM_SAT_O2^MDC.
Prescription level
Mandatory / Required / Shall
Page
165
Section
E.1.1 Example of PCD-01 Observation Report (Physiological Monitor)
Status
to be reviewed
Last changed
3/11/16 7:49:00 AM by mpattillo
Comment

Covered by

Tests

Test Steps

Rules

MBV

Covered by (Deprecated)

Test Steps

Rules

Document name
Provenance
Revision
Action
IHE Patient Care Device (PCD) Technical Framework - Volume 2IHE5.0