Assertion

AssertionId
LAW-005 Testable
predicate
The AWOS ID is never entered manually on the Analyzer. It can only be obtained via a message from the Analyzer Manager. Therefore, when the AWOS is entered at the Analyzer, the AWOS shall have a value of NULL ("") for the null AWOS ID.
Prescription level
Mandatory / Required / Shall
Page
36
Section
X.2.3
Status
validated
Last changed
3/7/16 2:39:37 PM by aberge
Comment

Applies to

Actors

AIPOs

Audit Messages

Integration Profiles

Transactions

Standards

Covered by

Tests

Test Steps

Rules

MBV

Covered by (Deprecated)

Tests

Test Steps

Rules

Document name
Provenance
Revision
Action
IHE PaLM Technical Framework Supplement - Laboratory Analytical Workflow (LAW)IHE1.5