Assertion

AssertionId
RAD68-031 Testable
predicate
XDSDocumentEntry Attribute: referenceIdList can describe both the Accession Number and Study Instance UID. The Accession Number value may be found in the Originating HL7 Imaging Order Message (OMI) Imaging Procedure Control (IPC) Segment, Sequence 1 for Accession Identifier, or in the DICOM attribute Accession Number (0008,0050). The referendIdList may be populated with the Accession Number and assigning authority. Only the CXi.1, CXi.4 and CXi.5 components shall be valued. The referencedIdList may contain multiple entries for Accession Number. The Study Instance UID uniquely identifies an imaging study. Only the CXi.1 and CXi.5 components shall be valued: CXi.1 - shall be the Study Instance UID CXi.5 shall be urn:ihe:iti:xds:2016:studyInstanceUID.
Prescription level
Permitted / May / Can
Page
477
Section
4.68.4.1.2.3.2
Status
reviewed
Last changed
2/1/24 9:55:02 AM by alegrand
Comment

Applies to

Audit Messages

Integration Profiles

Transactions

Standards

Covered by

Tests

Test Steps

Rules

MBV

Covered by (Deprecated)

Rules

Document name
Provenance
Revision
Action
IHE Radiology (RAD) Technical Framework Volume 206/15/2023