LAB27LAB27-001validatedTestable 0 5 After the Analyzer working in query mode recognizes one or more specimens, the Analyzer sends a WOS Query Message (QBP^Q11^QBP_Q11) for each specimen to the Analyzer Manager.168Section 3.Q.53/7/16 2:23:29 PM by mpattillo
LAB27LAB27-007validatedTestable 0 1 An analyzer supporting the "Query by Tray" Option SHALL be capable of sending a query identifying the position of the specimen in the tray. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-6, QPD-7 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-6 identifies the tray and QPD-7 is the position in the tray171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB28LAB28-008validatedTestable 0 5 A Negative Query Response for an unrecognized specimen is considered an unexpected situation, the Analyzer SHALL respond with an Application Acknowledgement: Reject in the ORL^O34 by setting MSA-1 to AR and reporting the appropriate specimen container values in the ERR segment.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo